Spectrum correction considering light source fluctuation for non-invasive blood glucose sensing

Thumbnail Image

Type:

Article

Date

2014-03-04

Journal Title

Journal ISSN

Volume Title

Publisher

SPIE

Research Projects

Organizational Units

Journal Issue

Abstract

The purpose of this study is to correct baseline shift in absorbance spectrums caused by light source fluctuation. To improve quantitative evaluation performance of blood glucose level, baseline shift is corrected by multiple scatter correction (MSC). Moreover, to increase the effect of the MSC, water vapor absorbance is subtracted, and relative glucose absorbance are calculated by dividing with hemoglobin absorbance at 1544 [cm-1]. In order to verify the effectiveness of the proposed spectrum correction method, light source fluctuation is simulated on the Fourier transform infrared spectroscopy (FT-IR), and we apply the proposed method to the spectrums measured by FT-IR. From the simulation results, the baseline shift was successfully reduced by proposed method.